K shell absorption jump ratios and jump factor measurements for some elements in the range of 40 <= Z <= 50

Yükleniyor...
Küçük Resim

Tarih

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

CANADIAN JOURNAL OF PHYSICS

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

In this study, K shell absorption jump ratios (J(K) ) and jump factors (r(K)) of specimens in the atomic number range of 40 <= Z <= 50 were measured using the energy dispersive X-ray fluorescence (EDXRF) technique. Related specimens were excited using 59.54 keV gamma-photons that were radiated from a Am-241 point source. Typical K X-rays emitting from specimens were detected from silicon drifted lithium (Si(Li)) semiconductor detector. Derived J(K) and r(K) values of related specimens were compared with calculated theoretical and other experimental values. The agreement of measured values was reasonable with theoretical calculations.

Açıklama

Anahtar Kelimeler

jump ratio, jump factor, K shell intensity ratio, X-ray, EDXRF

Kaynak

CANADIAN JOURNAL OF PHYSICS

WoS Q Değeri

Scopus Q Değeri

Cilt

97

Sayı

7

Künye

TURŞUCU, A. (2019). K shell absorption jump ratios and jump factor measurements for some elements in the range of 40 iZ i 50 title. Canadian Journal of Physics, 97(7), 786–790.